5.3.3 DOSIMATS

746 VA Trace Analyzer / 747 VA Stand

5-39

Softkeys

Meaning

Start

dilution

Start dilution operation

With this softkey the dilution operation for preparing standard solu-
tions at the selected Dosimat/Dosino is started. For this operation, the
6.5611.010 pipetting equipment must have been installed at the
Dosimat/Dosino and connected to the remote interface of the 746 VA
Trace Analyzer (see section 3.10.6).

To prepare a solution with the volume

V.end dilution

and the

concentration

C.end dilution

out of a standard with the concentra-

tion

C.init std

proceed as follows:

Select the Dosimat/Dosino with the mounted pipetting equipment
and press the softkey

for initializing the Dosi-

mat/Dosino and fill the transfer tubing.

Press the softkey

. The message

Preparing

Dosimat #X

appears.

Put the pipetting tip into a waste container at about the same level
as the container with the standard solution and press the button on
the pipetting tip. After aspiration of the dilution solution the mes-
sage

Aspirating standard

appears.

Put the pipetting tip into the standard solution and press the button
on the pipetting tip. After aspiration of the standard solution the
message

Dispensing solution

appears.

Put the pipetting tip into the desired void container and press the
button on the pipetting tip. Standard and dilution solution are dis-
pensed into the void container. Afterwards the message

Next

operation / press to break

appears.

If you press , the Dosimat/Dosino will be ready for
another dilution, if you press , the Dosimat/Dosino is
switched to the exchange position.

Exchange

all

Switch all Dosimats/Dosinos to exchange position

In order to change the exchange or dosing unit, the 685 Dosimats or
700 Dosinos must be in the

exchange

position. If the Dosi-

mats/Dosinos are in the

dosing

position, they have to be switched to

the exchange position by pressing

or

.

After the exchange or dosing units have been changed, the Dosi-
mats/Dosinos must be initialized by pressing

or

.

Prime

all

Initialize all Dosimats/Dosinos

The burette cylinders of all Dosimats/Dosinos and all connecting
tubing up to the measuring vessel are automatically filled. Excess
solution is expelled to the measuring vessel.
Should the filling take longer than 30 s, the message

This operation

takes x.x minutes

appears. The filling operation can then be started

with or aborted with to increase the filling rate. Once a filling
operation has been started, it can no longer be aborted.
After the filling, the initial volume

V.init

is automatically delivered to

the measuring vessel. With

V.init > 0 uL

, the Dosimat/Dosino is

then in the

dosing

position and is ready for dispensing. With

V.init

= 0 uL

, the

exchange

position is retained, the Dosimat/Dosino is

ready for both dispensing and replacement of the exchange or dosing
unit.