5.6.3 SUBSTANCES

746 VA Trace Analyzer / 747 VA Stand

5-151

Parameter

Range; Default value / Meaning

M.conc./cm

0.001...999.99 ng...kg/L;

Axis scaling for plot of calibration curve

Here, the scale in g/L per cm for the plot of the x axis of the calibration
curve (

ρ

eff

) can be entered. In the absence of an entry, the entire curve

is plotted within 10 cm.

M.conc.begin

auto, zero; auto

Begin of x axis for plot of calibration curve

Selection of the begin of the x axis (mass concentration) for the plot of
the calibration curve:

auto

Automatic selection of the begin

zero

Begin fixed to 0 g/L

V0.CSoln.

0.1...50.0 mL; 5.000 mL

Addition volume of calibration solutions

Volume of every calibration solution that must be added to the meas-
uring vessel.
If work is performed with the 695 Autosampler, an asterisk

*

can be

entered here in place of a volume. However, this is admissible only if
the instruction

CSOLN/F

with subsequent solution transfer to the

auxiliary or measuring vessel is used.
If calibration curves of several substances are recorded simultane-
ously,

V0.CSoln.

must be the same for every substance.

Coefficients

display only

Coefficients

Title for coefficients of the calibration curve.

Y.offset

-999.9...999.9 e-20...e+20;

Intercept of calibration curve

Display of the intercept of the calculated calibration function of type

linear

or

nonlin

in A, W or s. This value can be changed manually.

This value can be used in formulae as

KY:

(

= sub-

stance name) for further calculations.

Slope

-999.9...999.9 e-20...e+20;

Slope of calibration curve

Display of the slope of the calculated calibration function in AL/g,
WL/g or sL/g. This value can be changed manually.

With non-linear curves,

Slope

represents the slope in the linear region

(see also

Curve type

).

This value can be used in formulae as

KS:

(

= sub-

stance name) for further calculations.

Nonlin.

-999.9...999.9 e-20...e+20;

Non-linearity factor

Display of the non-linearity factor of the calculated nonlinear calibra-
tion function in [Al/g]

4

, [WL/g]

4

or [sL/g]

4

(see also

Curve type

). This

value can be changed manually.

This value can be used in formulae as

KN:

(

= sub-

stance name) for further calculations.