Metrohm 746 VA Trace Analyzer User Manual
Page 302

5.6.3 SUBSTANCES
746 VA Trace Analyzer / 747 VA Stand
5-151
Parameter
Range; Default value / Meaning
M.conc./cm
0.001...999.99 ng...kg/L;
Axis scaling for plot of calibration curve
Here, the scale in g/L per cm for the plot of the x axis of the calibration
curve (
ρ
eff
) can be entered. In the absence of an entry, the entire curve
is plotted within 10 cm.
M.conc.begin
auto, zero; auto
Begin of x axis for plot of calibration curve
Selection of the begin of the x axis (mass concentration) for the plot of
the calibration curve:
auto
Automatic selection of the begin
zero
Begin fixed to 0 g/L
V0.CSoln.
0.1...50.0 mL; 5.000 mL
Addition volume of calibration solutions
Volume of every calibration solution that must be added to the meas-
uring vessel.
If work is performed with the 695 Autosampler, an asterisk
*
can be
entered here in place of a volume. However, this is admissible only if
the instruction
CSOLN/F
with subsequent solution transfer to the
auxiliary or measuring vessel is used.
If calibration curves of several substances are recorded simultane-
ously,
V0.CSoln.
must be the same for every substance.
Coefficients
display only
Coefficients
Title for coefficients of the calibration curve.
Y.offset
-999.9...999.9 e-20...e+20;
Intercept of calibration curve
Display of the intercept of the calculated calibration function of type
linear
or
nonlin
in A, W or s. This value can be changed manually.
This value can be used in formulae as
KY:
(
= sub-
stance name) for further calculations.
Slope
-999.9...999.9 e-20...e+20;
Slope of calibration curve
Display of the slope of the calculated calibration function in AL/g,
WL/g or sL/g. This value can be changed manually.
With non-linear curves,
Slope
represents the slope in the linear region
(see also
Curve type
).
This value can be used in formulae as
KS:
(
= sub-
stance name) for further calculations.
Nonlin.
-999.9...999.9 e-20...e+20;
Non-linearity factor
Display of the non-linearity factor of the calculated nonlinear calibra-
tion function in [Al/g]
4
, [WL/g]
4
or [sL/g]
4
(see also
Curve type
). This
value can be changed manually.
This value can be used in formulae as
KN:
(
= sub-
stance name) for further calculations.