beautypg.com

276 scanctrl, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual

Page 427

background image

5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-276

SCANCTRL

Page:

O, S

Stand:

M, A

Abbreviation:

S

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. SCANCTRL ..... Code ........ Test ....

Function

Inquiry of the input control lines 1...8 and stop of the program until either
the test condition is met or the line time has elapsed.

Parameter

Range; Default value / Meaning

Code

1, 0, *; *

Comparison code for test

8-digit numeric code for status of the input control lines 1...8. The
following values can be entered for each line:

0

off, inactive (5 V)

1

on, active

(0 V)

*

immaterial

Test

=, <>; =

Condition for test

This parameter can be used to select whether the program run should
be continued on correspondence or non-correspondence of compari-
son status and actual status of the input control lines:

=

Program continuation if status of the control lines =
comparison code (or when line time has elapsed)

<>

Program continuation if status of the control lines

comparison code (or when line time has elapsed)

Description

The

SCANCTRL

instruction is used to stop the program run and poll the

status of the input control lines 1...8 of the remote interface "Control
Lines". The program is not continued until its status is equal (

=

) or not

equal (

<>

) to the preset comparison status

Code

or after elapse of the

line time.

The

SCANCTRL

instruction can be used to monitor functions of external

devices attached to the remote interface.

Comments

The attachment of devices to the remote interface is described in section

3.13.

see also

SETCTRL