5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-224

The current measurement with the

DCTMODE

measurement mode

normally provides wave-shaped curves which can be evaluated using
the tangent method (baseline of type

linear

) (see section 5.6.3).

U.wave

I.wave

I

U

If peak-shaped curves appear with the

DCTMODE

mode, these can be

evaluated only if the voltage sweep

SSWEEP

is used. The evaluation is

then the same as with other peaks using the baselines of type

linear

,

polynom

or

expon.

(see section 5.6.3).

I

U

U.peak

I.peak

Comments

The

DCTMODE

measurement mode can be used with all types of

electrodes together with the instructions

MEAS

,

SWEEP

,

DSWEEP

,

FSWEEP

or

TSWEEP

.

The electrode type

DME

,

SMDE

,

HMDE

or

RDE

must be entered before

the

DCTMODE

measurement mode.

With the

DCTMODE

measurement mode, the potentiostat always

operates in the operating mode "fast".

SWEEP

and

FSWEEP

are thus

identical for this mode.

The quantity

t.drop

important for the definition of

t.step

depends

on the type of electrode selected:

DME

t.drop = 20 ms

HMDE and RDE

t.drop = 0 ms

SMDE

t.drop = Drop.size

×× 40 ms

see also

SWEEP, SSWEEP, DME, HMDE, SMDE, RDE