5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-224
The current measurement with the
DCTMODE
measurement mode
normally provides wave-shaped curves which can be evaluated using
the tangent method (baseline of type
linear
) (see section 5.6.3).
U.wave
I.wave
I
U
If peak-shaped curves appear with the
DCTMODE
mode, these can be
evaluated only if the voltage sweep
SSWEEP
is used. The evaluation is
then the same as with other peaks using the baselines of type
linear
,
polynom
or
expon.
(see section 5.6.3).
I
U
U.peak
I.peak
Comments
•
The
DCTMODE
measurement mode can be used with all types of
electrodes together with the instructions
MEAS
,
SWEEP
,
DSWEEP
,
FSWEEP
or
TSWEEP
.
•
The electrode type
DME
,
SMDE
,
HMDE
or
RDE
must be entered before
the
DCTMODE
measurement mode.
•
With the
DCTMODE
measurement mode, the potentiostat always
operates in the operating mode "fast".
SWEEP
and
FSWEEP
are thus
identical for this mode.
•
The quantity
t.drop
important for the definition of
t.step
depends
on the type of electrode selected:
DME
t.drop = 20 ms
HMDE and RDE
t.drop = 0 ms
SMDE
t.drop = Drop.size
×× 40 ms
see also
SWEEP, SSWEEP, DME, HMDE, SMDE, RDE