Metrohm 746 VA Trace Analyzer User Manual
Page 147

4 Introduction to the operation
746 VA Trace Analyzer / 747 VA Stand
4-62
4.6.3
Method specifications of the method
ASV.mth
In this section the 5 dialog pages of the group "Method Specifications" associated
with the loaded method
ASV.mth
will be explained in detail. If you are interested
only in a rapid lead measurement, you can proceed directly to section 4.6.4.
1
Switch to dialog page "OPERATION SEQUENCE"
•
Press the yellow
SEQUENCE" appears with the main program:
Method: ASV .mth OPERATION SEQUENCE
Title : Anodic stripping voltammetry / -1150 mV -> 150 mV
Instructions t/s Main parameters Auxiliary parameters
ДДДДДДДДДДДДД
ДДДДДДДДДДДДД ДДДДД
ДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДД
ДДДДДДДДДДДДДДДДДДДДДДДДД
ДДДДДДДДДДДДДДДДДДДДДДДДД
ДДДДДДДДДДДДДДДДДДДДДДДДД
1 SMPL>M V.fraction mL V.total L
2 DOS>M Soln.name Acetate V.add 1.000 mL
3 PURGE
4 STIR 300.0 Rot.speed 2000 /min
5 ! (ADD
6 PURGE
7 STIR 30.0 Rot.speed 2000 /min
8 ØPURGE
9 (REP
10 SEGMENT Segm.name ASV
11 REP)0
12 ! ADD>M Soln.name V.add 0.100 mL
13 ! ADD)0
14 END
Modify
column
Modify
line
Delete
instr.
Insert
instr.
Method
labels
•
The following is a brief explanation of the main program sequence:
Line 1
Addition of 10 mL sample
Line 2
Addition of 1 mL acetate buffer 1mol/L pH 4.6
Line 3....4
Purging and stirring of the analysis solution for 5 min
Line 5...13
Prepared loop for standard addition (inactive)
Line 6....7
Purging and stirring of the analysis solution for 30 s
Line 8
Switching off the purging
Line 9...11
Prepared loop for replication of the segment
Line 10
Call-up of the segment
ASV
Line 12
Prepared instruction for addition of the standard
addition solution (inactive)
Line 14
End of main program