4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-62

4.6.3

Method specifications of the method

ASV.mth

In this section the 5 dialog pages of the group "Method Specifications" associated

with the loaded method

ASV.mth

will be explained in detail. If you are interested

only in a rapid lead measurement, you can proceed directly to section 4.6.4.

1

Switch to dialog page "OPERATION SEQUENCE"

Press the yellow key. The dialog page "OPERATION

SEQUENCE" appears with the main program:

Method: ASV .mth OPERATION SEQUENCE

Title : Anodic stripping voltammetry / -1150 mV -> 150 mV

Instructions t/s Main parameters Auxiliary parameters

ДДДДДДДДДДДДД

ДДДДДДДДДДДДД ДДДДД

ДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД

1 SMPL>M V.fraction mL V.total L

2 DOS>M Soln.name Acetate V.add 1.000 mL

3 PURGE

4 STIR 300.0 Rot.speed 2000 /min

5 ! (ADD

6 PURGE

7 STIR 30.0 Rot.speed 2000 /min

8 ØPURGE

9 (REP

10 SEGMENT Segm.name ASV

11 REP)0

12 ! ADD>M Soln.name V.add 0.100 mL

13 ! ADD)0

14 END

Modify

column

Modify

line

Delete

instr.

Insert

instr.

Method

labels

The following is a brief explanation of the main program sequence:

Line 1

Addition of 10 mL sample

Line 2

Addition of 1 mL acetate buffer 1mol/L pH 4.6

Line 3....4

Purging and stirring of the analysis solution for 5 min

Line 5...13

Prepared loop for standard addition (inactive)

Line 6....7

Purging and stirring of the analysis solution for 30 s

Line 8

Switching off the purging

Line 9...11

Prepared loop for replication of the segment

Line 10

Call-up of the segment

ASV

Line 12

Prepared instruction for addition of the standard
addition solution (inactive)

Line 14

End of main program