4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-8

747 VA STAND

Manual Address 1

Switches Settings States

ДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД

Pressure off U.standby --- mV

Purging off

Stirrer off Rot.speed 2000 /min

DME off

HMDE off Drop size 5

SMDE off Drop size 5

RDE off Rot.speed 2000 /min

V.meas max 80 mL

VA Stand

1

VA Stand

2

695 Auto

sampler

717 Smpl

Changer

Rinsing

Dosinos

Electr.

test

Default

values

>>

Switches

8

Electrode test of the DME

Press the softkey

. The four softkeys

,

,

and

appear in the softkey bar.

Press the softkey

. The message

Connect Dummy Cell or

electrodes

appears.

Press to start the test. The DME is tested, the message

Testing electrodes

appears. On successful completion of the test, the

message

DME test ok

appears in the system window. Press the

key to close the system window.

If the DME test was not successful, an error message appears (see

section 7.3.2). Check the installation of the MME as described in section

3.4 and then repeat the test.

You will find further information regarding the DME under the instruction DME in section
5.7.3.

9

Electrode test of the HMDE

Press the softkey

. The four softkeys

,

,

and

appear in the softkey bar.

Press the softkey

. The message

Connect Dummy Cell or

electrodes

appears.

Press to start the test. The HMDE is tested, the message

Testing electrodes

appears. On successful completion of the test, the

message

HMDE test ok

appears in the system window. Press the

key to close the system window.

If the HMDE test was not successful, an error message appears (see

section 7.3.2). Check the installation of the MME as described in section

3.4 and then repeat the test.

You will find further information regarding the HMDE under the instruction HMDE in
section 5.7.3.