5.5 Operation

746 VA Trace Analyzer / 747 VA Stand

5-86

Parameter

Range; Default value / Meaning

In evaluations with standard addition (see section 6.9.4),

MC.dev.

is

calculated directly from the measured values of the sample and
standard additions. In evaluations with a calibration curve (see section

6.9.5) or internal standard (see section 6.9.6),

MC.dev.

comprises the

deviation component of the calibration

Cal.dev.

and that of the

measurement

Meas.dev.

as follows:

MC.dev. = (Cal.dev.)

2

+ (Meas.dev.)

2

Cal.dev.

display only

Deviation due to calibration (variable ML:)

Deviation of the mass concentration

ρ

s

of the substance due to the

calibration (calibration curve or internal standard) (see also

MC.dev.

).

The absolute value with numeric value and unit together with, in
brackets, the relative value in % is shown (for further details, see

section 6.9).
The deviation of the mass concentration

ρ

s

of the substance due to

the calibration can be used in formulae as

ML:

(

=

substance name) for further calculations. It should be noted that this
value is put out in such cases in the base unit g/L without prefix (see

section 5.6.4).
This field remains empty (

) if standard addition has been selected as

the calibration technique (

Technique = std.add.

) or no result has

been found.

Mass

display only

Substance mass in measuring vessel (variable MS:)

Absolute mass of the selected substance in the measuring vessel with
numeric value and unit (g, mg,

µ

g, etc.) determined with the aid of the

selected calibration (

std.add.

,

cal.crv.

,

int.std.

).

The absolute mass of the substance in the measuring vessel can be
used in formulae as

MS:

(

= substance name) for

further calculations. It should be noted that this value is put out in
such cases in the base unit g without prefix (see section 5.6.4).
The relation between

Mass

and

Mass conc.

is given by

Mass conc. = Mass / V0.sample

This field remains empty (

-

) if no calibration technique has been

selected (

Technique = none

) or no result has been found.

Add.mass

display only

Standard addition mass (variable MA:)

Mass of the selected substance with numeric value and unit (g, mg,

µ

g, etc.) added in every standard addition.

The standard addition mass can be used in formulae as

MA:

(

= substance name) for further calculations. It should be

noted that this value is put out in such cases in the base unit g without
prefix (see section 5.6.4).
This value can be used to asses whether the standard addition ratio
has been correctly selected (cf. remarks on

std.add.

in section

5.6.3).
This field remains empty (

) if standard addition has not been se-

lected as the calibration technique (

Technique

≠≠ std.add.

).