beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 281

background image

5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-130

Softkeys

Meaning

Modify

column

Modify parameters column-by-column

Modify

line

Modify parameters line-by-line

Select

subst.

Select substance

The existing substances and those still free can be selected with the
softkeys:

Subst.#

Segment#

Selection of a substance

Subst.#

already linked to the segment

Segment#

Subst.#

Selection of a substance

Subst.#

not linked to a segment

--------

Selection of a substance still free. The substance is automatically
given the standard name

Subst#

, which can later be changed with the

softkey

.

Link to

segment

Link substance to segment

Substances can be evaluated only if they are linked to a predefined
segment. This is accomplished using the softkeys which display all
segments defined on the "SEGMENTS" page.

Segment#

Linkage of the selected substance to the segment

Segment#

. Follow-

ing the linkage, the segment name is shown with the substance name
in the title line.

Delete

subst.

Delete substance

The selected substance will be deleted directly without any warning.

Subst.

name

Change substance name

Possibility to change the substance name

Subst#

allocated as

standard.

For the substance name, no brackets "(" or ")" are allowed.

>>

Switching to subpage "Calibration"

Switching to the subpage "Calibration" with details of the substance
calibration.

Keys

Meaning

ENTER

Substance parameter test

By pressing the key in the navigation mode the parame-
ters on the "SUBSTANCES" page are tested:

If no error is found, no action occurs.

If errors are found, the appropriate error messages are displayed
in the error window (see section 7.3.2).

21