Metrohm 746 VA Trace Analyzer User Manual
Page 281

5.6 Method Specifications
746 VA Trace Analyzer / 747 VA Stand
5-130
Softkeys
Meaning
Modify
column
Modify parameters column-by-column
Modify
line
Modify parameters line-by-line
Select
subst.
Select substance
The existing substances and those still free can be selected with the
softkeys:
Subst.#
Segment#
Selection of a substance
Subst.#
already linked to the segment
Segment#
Subst.#
Selection of a substance
Subst.#
not linked to a segment
--------
Selection of a substance still free. The substance is automatically
given the standard name
Subst#
, which can later be changed with the
softkey
.
Link to
segment
Link substance to segment
Substances can be evaluated only if they are linked to a predefined
segment. This is accomplished using the softkeys which display all
segments defined on the "SEGMENTS" page.
Segment#
Linkage of the selected substance to the segment
Segment#
. Follow-
ing the linkage, the segment name is shown with the substance name
in the title line.
Delete
subst.
Delete substance
The selected substance will be deleted directly without any warning.
Subst.
name
Change substance name
Possibility to change the substance name
Subst#
allocated as
standard.
For the substance name, no brackets "(" or ")" are allowed.
>>
Switching to subpage "Calibration"
Switching to the subpage "Calibration" with details of the substance
calibration.
Keys
Meaning
ENTER
Substance parameter test
By pressing the
ters on the "SUBSTANCES" page are tested:
•
If no error is found, no action occurs.
•
If errors are found, the appropriate error messages are displayed
in the error window (see section 7.3.2).
21