5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-296

Description

The

SSWEEP

instruction starts the voltage sweep defined in the parame-

ters.

U

t

U.start

U.end

U.step

t.step

Comments

The electrode type must be defined before the

SSWEEP

instruction.

Admissible are

DME

,

SMDE

,

HMDE

or

RDE

.

The measurement mode must be defined before the

SSWEEP

instruction. Admissible is only

DCTMODE

.

Apart from the

SSWEEP

instruction, no other SWEEP instruction is

allowed in the same segment.

If the final voltage

U.end

should not be switched off at the end of the

program, enter a suitable standby voltage

U.standby

in a

ØMEAS

instruction.

With the

SSWEEP

instruction maximum 1000 measured values can be

recorded.

see also

SWEEP, DSWEEP, FSWEEP, DME, SMDE, HMDE, RDE, DCTMODE