5.4.4 SAMPLES

746 VA Trace Analyzer / 747 VA Stand

5-61

Parameter

Range; Default value / Meaning

Name

8 characters;

Name of sample data table

Name of the sample data table which is used as file name together
with the automatic extension

.smp

in storage of the sample data on

the data card or RS interface. The softkey

must be

pressed to change the name.

For the name of the sample table all characters are allowed with the

exception of the character "

.

".

Line

display only

Line number

Number of the sample data line on which the field cursor is posi-
tioned.

Autoincrement

1,2; 1

Increment for automatic position numbering

Increment for the automatic numbering of the

Pos.

column during

filling in of the sample data table (effective if a position number is
entered in the previous line):

1

continuous increase in the position number by +1

2

continuous increase in the position number by +2
(useful if an auxiliary vessel AUX is used after every
sample vessel FEED on the Autosampler).

The parameter

Autoincrement

is changed with

.

Balance input

on, off; off

Data receipt from balance

The data receipt from a balance attached to RS232 interface 2 can be
switched on or off using the softkey

. When the data

receipt is switched on, the data sent by the balance are written into
the next blank line.

For the attachment of balances, see section 3.12.3.

Pos.

1...160;

Position number

Position number on the sample rack of the 695 Autosampler (without
695 Autosampler this column has no meaning, however it can be
used for simple sample numbering).

The following rules apply to the position numbering:

All instructions concerning the sample vessel (FEED) (e.g.

SMPL/F

,

CSOLN/F

,

F>M

,

DOS>F

, etc.) are referred to the vessel at

the rack position

Pos.

defined by the position number.

All instructions concerning the auxiliary vessel (AUX) (e.g.

F>A

,

DOS>A

, etc.) are referred to the vessel which immediately follows

the sample vessel defined by the position number (rack position

Pos. + 1

).

The sequence of the positions can be freely selected for sample
solutions, but the same position number can be entered only
once.