beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 89

background image

4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-4

746 VA TRACE ANALYZER

Program 5.746.0101

Date 1997-03-20 RS232 settings Ifc.1 Ifc.2

Time 10:41:15 ––––––––––––––––––––––––––––––––––––––––––––

Beeper on Baud rate 9600 9600

Data bits 8 8

Display Stop bit(s) 1 1

––––––––––––––––––––––––––––– Parity none none

Switch–off time 30 min Handshake Hwshort Hwshort

Contrast 106 Receive mode off off

Configuration Address Balance type none

––––––––––––––––––––––––––––– Printer type none none

VA stand 1 Printer width 1.00 1.00

Remote keyboard off

Ctrl.lines # 87654321 Printer

––––––––––––––––––––––– ––––––––––––––––––

Scan 00000000 Type Metrohm

Set 00000000

Ext. Start off

Modify

column

Set Ctrl

lines

Contrast

up

Contrast

down

Reset

Ifc.1

Reset

Ifc.2

Save

Ifc.par.

Diagn.

Dialog pages are divided into three parts:

− The header contains the page title and further details regarding the dialog page (here the

program number).

− The largest part of the dialog page is taken up with the page parameters. All editable pa-

rameters have not only a parameter name but also a gray-backed entry field in which a nu-
meric value or text expression can be entered. The individual entry fields can be selected in the
navigation mode (see section 5.2.2) using the cursor keys (arrow keys) with the current posi-
tion of the field cursor being marked by the black-backed entry field.

− The bottom part of each dialog page contains the softkey bar in which the functions for the 8

softkeys F1...F8 are displayed.

You will find further information regarding the layout of the dialog pages in section 5.2.1.

3

Setting the date

After selection of the "VA TRACE ANALYZER" dialog page the field cursor

is already located in the entry field

Date

, which is black-backed.

If you wish to change the displayed date, first press the space bar to

change from the navigation mode to the edit mode. The whole field is

now no longer black-backed, but only the first flashing digit, at the same

time all displays in the softkey bar except for the softkey function

have disappeared.

Entry into the edit mode to change a parameter can be effected not only with the
space bar, but also with the key, with the softkeys

and as well as by pressing all alphanumeric keys which are not in
simultaneous use as dialog keys (for details, see section 5.2.3).

Press the softkey

. A help message for the

Date

parameter

appears in the

HELP WINDOW

with details of the entry format and the pos-

sible entry values. Close this window by pressing the key

again.