beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 277

background image

5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-126

Parameter

Range; Default value / Meaning

##

display only

Line number

Automatic numbering of the instruction lines (max. 29 instruction lines
can be entered).

Attribute

C, D, !, @;

Line attribute

Optional marking of the instruction lines with the following attributes:

C

Calibration; the marked instruction line is executed only
if

Run mode = calibration

.

D

Determination; the marked instruction line is executed
only if

Run mode = determination

.

!

Inactive line; this line will not be executed.

@..@

Inactive range; the range between the two lines marked
by

@

will not be executed.

Combinations of line attributes are also possible (e.g.

C!

,

D@

).

Instructions

Instructions;

Program instructions

Entry of program instructions (see section 5.7). The associated
parameters are entered automatically.

In the entry, all possible instructions on this dialog page are displayed
in alphabetical order on the softkeys and can thus be incorporated
directly:

XXXXXXXX

--------

Instruction

XXXXXXXX

is copied to the

Instructions

field.

t/s

0.1...999.9 s;

Line time

Optional entry of a wait time for the instruction line. When the program
is running, this line time will be allowed to elapse before the next
instruction is executed. The time counter is started at the moment the
program instruction is started.

With SWEEP instructions, the line time is calculated automatically and
can not be changed.

Main parameters

Parameter;

Main parameters

Main parameters for program instruction (see section 5.7.3).

Auxiliary
parameters

Parameter;

Auxiliary parameters

Auxiliary parameters for program instruction (see section 5.7.3).