5.6 Method Specifications
746 VA Trace Analyzer / 747 VA Stand
5-126
Parameter
Range; Default value / Meaning
##
display only
Line number
Automatic numbering of the instruction lines (max. 29 instruction lines
can be entered).
Attribute
C, D, !, @;
Line attribute
Optional marking of the instruction lines with the following attributes:
C
Calibration; the marked instruction line is executed only
if
Run mode = calibration
.
D
Determination; the marked instruction line is executed
only if
Run mode = determination
.
!
Inactive line; this line will not be executed.
@..@
Inactive range; the range between the two lines marked
by
@
will not be executed.
Combinations of line attributes are also possible (e.g.
C!
,
D@
).
Instructions
Instructions;
Program instructions
Entry of program instructions (see section 5.7). The associated
parameters are entered automatically.
In the entry, all possible instructions on this dialog page are displayed
in alphabetical order on the softkeys and can thus be incorporated
directly:
XXXXXXXX
--------
Instruction
XXXXXXXX
is copied to the
Instructions
field.
t/s
0.1...999.9 s;
Line time
Optional entry of a wait time for the instruction line. When the program
is running, this line time will be allowed to elapse before the next
instruction is executed. The time counter is started at the moment the
program instruction is started.
With SWEEP instructions, the line time is calculated automatically and
can not be changed.
Main parameters
Parameter;
Main parameters
Main parameters for program instruction (see section 5.7.3).
Auxiliary
parameters
Parameter;
Auxiliary parameters
Auxiliary parameters for program instruction (see section 5.7.3).