5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-201

(CAL

Page:

O, S

Stand:

M, A

Abbreviation:

(C

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. (CAL .....

Function

Start of the loop for the recording of a calibration curve.

Description

The

(CAL

instruction opens a loop for recording a calibration curve with

both the calibration technique with calibration curve

cal.crv.

and that

with the internal standard

int.std.

. The CAL loop must be closed with

the

CAL)n

instruction.

In calibrations with a calibration curve, for the recording of this curve

the CAL loop must include one of the dosing instructions

CSOLN/M

,

CSOLN>M

or

CSOLN/F

for addition of the calibration solutions as well as a

sweep or segment call-up. In addition, any other instructions can be
incorporated.

In calibrations with an internal standard, for the recording of the

calibration curve for the determination of the proportionality factor the
CAL loop must include one of the dosing instructions

CSOLN/M

,

CSOLN>M

or

CSOLN/F

for addition of the calibration solutions, the instruction

RADD>M

or

RADD>F

for addition of the internal standard as well as a

sweep or segment call-up. In addition, any other instructions can be
incorporated.

Comments

A program can contain several CAL loops, but they must not be
nested.

REP loops are allowed within CAL loops, a CAL loop can also be
inserted within a REP loop.

No ADD or VAR loops are allowed within CAL loops.

It is advantageous to insert the

ØMEAS

instruction in the CAL loop to

switch off the electrode voltage when the solution is changed.

If the entire analysis solution is changed each time during recording
of the calibration curve, a

RINSE

instruction must be incorporated in

the CAL loop which resets the volume bookkeeping in the measuring
vessel to 0 (see also

RINSE). If its parameter

Cycles > 0

is set, the

measuring vessel is also rinsed by means of the two rinsing Dosinos
connected to the 746 VA Trace Analyzer or the 695 Autosampler. If no
rinsing Dosinos are connected, the prompt

Change measuring

vessel

to change the solution appears.

Without a

RINSE

instruction, the volume in the measuring vessel is

added continuously. This can be utilized for the recording of calibra-
tion curves in which the calibration solutions are prepared each time
by manual addition of a concentrated standard solution to the meas-
uring vessel (see

CSOLN>M).

The number of calibration solutions is defined in the

CAL)n

instruc-

tion. A valid mass concentration

Mass conc.

must be entered on the

"SUBSTANCES/Calibration" page for each of these solutions.