5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-280

SEGMENT

Page:

O

Stand:

M, A

Abbreviation:

SE

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. SEGMENT Segm.name ........

Function

Call-up of a segment.

Parameter

Range; Default value / Meaning

Segm.name

8 characters; -

Segment name

The name entered here must be identical to the name of the segment
which should be executed.

Description

The

SEGMENT

instruction executes the segment with the name

Segm.name

and the instructions it contains.

Comments

A segment can be called up several times.

The call-up of a segment from another segment is not possible.