746 VA Trace Analyzer / 747 VA Stand

6-5

6.2.3

Method-independent settings

Depending on the method selected and before the start of a definition, in addition

to the parameters defined on the method dialog pages other method-independent

parameter settings are needed on other dialog pages:

MONITORING

(see section 5.5.1)

Run mode

Run mode (determination, calibration or test)

User name

Optional entry of the user name

Stand address

Entry of the stand address (1, 2 or 3)

Auto.samples

Automatic transfer of sample data from the

"SAMPLES" page

Auto.batch

Automatic series determination of the specified

number of samples

Det.storage

Automatic storage of determinations

Sample data

Pos.

,

Ident.1/S1

,

Ident.2/S2

,

Ident.3/S3

,

Method.call

,

Sample size/S0

(see also section

6.2.2)

SAMPLES

(see section 5.4.4)

Entry of sample data if these should be transferred automatically to the

"MONITORING" page (with

Auto.samples = on

).

DOSIMATS

(see section 5.3.3)

Setting of parameters for 685 Dosimats and 700 Dosinos used for the auto-

matic addition of standard addition or auxiliary solutions.

SOLUTIONS

(see section 5.3.4)

Definition of standard addition and auxiliary solutions which should be added

automatically with the 685 Dosimats and 700 Dosinos or with the 695 Auto-

sampler.
Before the start of a determination, ensure that all solutions defined in the

method for automatic dispensing are entered on this page and that these

solutions are located in the correct positions on the Dosimats or the Auto-

sampler.

COMMON VARIABLES (see section 5.4.5)

Definition of variables common to all methods.