Metrohm 746 VA Trace Analyzer User Manual
Page 124

4.3 Lead determination with calibration curve
746 VA Trace Analyzer / 747 VA Stand
4-39
4
Start the determination
•
Press the red
•
The status display in the header of the dialog page "MONITORING"
switches to
* BUSY *
, the start time of the measurement is entered un-
der
Determination
and the determination is given a standard name,
also formed from the start time. Further, during the measurement the
measurement window is superimposed which shows the current pro-
gram line being executed and the remaining line time.
•
Shortly after the start of the determination, the screen shows the prompt
to add 10 mL sample to the measuring vessel:
5
Addition of sample
•
Remove stopper 25
25 on 747 VA Stand and use a pipette to add 10 mL of
drinking water to the measuring vessel through the pipetting opening.
Press the
•
The screen shows the prompt to add 1 mL acetate buffer to the measur-
ing vessel:
6
Addition of acetate buffer
•
Use a pipette to add 1 mL acetate buffer (1 mol/L, pH 4.6) to the
measuring vessel through the pipetting opening. Then close the pipet-
ting opening using stopper 25
25 and press the
7
Measurement of the sample solution
•
Measurement of the sample solution runs as follows (see also section
4.3.3):
−
Purging and stirring for 180 s.
−
Call-up of the segment
DPASV
for measured value recording. The
lead present in the solution is enriched at the hanging mercury
drop with stirring at a measurement voltage of –600 mV and then
determined anodically without stirring from –600
…
–200 mV. The
selected measurement mode DP leads to a current peak whose
maximum lies at ca. –400 mV.
−
The
DPASV
segment is then repeated twice. The three single
measurements are allocated the VR numbers
00
,
01
and
02
.
•
On completion of the last measurement, the lead mass concentration is
calculated using the coefficients of the calibration curve stored in the
method. The program is now at an end. The status display in the header
of the dialog page "MONITORING" switches back to
* READY *
, the in-
strument is ready for a new determination.
Ъ
ЪДДД
ДДД MEAS.TASK MESSAGE ДДДДДДДДДДДДДДДДДД
ДДДДДДДДДДДДДДДДДД¿
¿
³
³
³
³
³
³
SMPL 10.000 mL > M
³
³
³
³
³
³
А
АДДД
ДДД
ДДДДДДДДДДДДДДДДДДЩ
Щ
Ъ
ЪДДД
ДДД MEAS.TASK MESSAGE ДДДДДДДДДДДДДДДДДД
ДДДДДДДДДДДДДДДДДД¿
¿
³
³
³
³
³
³
DOS Acetate 1.000 mL > M
press
³
³
³
³
³
³
А
АДДД
ДДД
ДДДДДДДДДДДДДДДДДДЩ
Щ