Metrohm 746 VA Trace Analyzer User Manual
Page 526

7.3 Error messages
746 VA Trace Analyzer / 747 VA Stand
7-27
Error message
Meaning/Causes
Rectification
Exit
SUBSTANCES - ERROR 3:
Param. Mass conc.
missing
Missing value for the
Mass
conc.
parameter
Enter
Mass conc.
parameter on the
”SUBSTANCES” page
Esc
SUBSTANCES - ERROR 4:
Param. Scope f./r.half
only with linear
The
f.half/r.half
pa-
rameter is admissible only with
a linear baseline
Change
Scope
or
Type
parameter on
the ”SUBSTANCES” page
Esc
SUBSTANCES - ERROR 5:
Param. Scope
f./r.double missing
counterpart
Missing counterpart for the
evaluation with
f.double
or
r.double
; in double peak
evaluation the first peak must
be evaluated with
f.double
and the second peak with
r.double
Check
Scope
parameter for both
substances and change
Esc
SUBSTANCES - ERROR 6:
Param. Scope
f./r.double direction
conflict
Wrong assignment of the two
peaks in the double peak
evaluation; the first peak (in
sweep direction) must be
evaluated with
f.double
, the
second with
r.double
Check
Scope
parameter for both
substances and change
Esc
SUBSTANCES - ERROR 7:
Undefined error
Non-specified error on the
”SUBSTANCES” page
Check the ”SUBSTANCES” page
Esc
VA STAND - ERROR 1:
Missing device
Connection to 747 VA Stand
broken:
−
Connecting cable not
plugged in
−
Wrong address
−
Hardware fault at VA Stand
Check connection to 747 VA Stand:
−
Plug in connecting cable
−
Correct address and restart 746 VA
Trace Analyzer
−
Run diagnostic routine for VA Stand
E/S
VA STAND - ERROR 2:
Working electrode
faulty
Fault at MME working electr.:
−
No connection
−
Too less mercury
−
Bad mercury quality
−
Dirty or clogged capillary
−
Dirty or damaged sealing
needle
−
No inert gas pressure
−
Dirty electrode body
−
Fault at reference electrode
−
No solution in measuring
vessel
−
No electrolyte in measuring
vessel
−
Tapping strength in DME
operation too weak
Fault at RDE working electrode:
−
No connection
−
Electrode tip not firmly
screwed on to the drive shaft
−
Contaminated electrode
surface
Check MME:
−
Connect reference electrode with WE
cable
−
Fill MME with mercury (see section
3.4.2 and 3.4.7)
−
Clean MME (see section 3.4.9) and
use ultrapure mercury (w
≥
0.99999)
or filtered mercury
−
Change capillary (see section 3.4.8)
−
Clean MME (see section 3.4.9) and
replace sealing needle if need be
−
Check inert gas connection (see
section 3.3.3)
−
Rub the MME body dry, clean with
distilled water of dilute acid and dry
with a lint-free cloth
−
see
VA STAND - ERROR 4
−
Add measuring solution
−
Add a few drops of KCl 3 mol/L to the
measuring solution
−
Turn slotted screw 71
71 on the 747 VA
Stand during operation in an anti-
clockwise direction until a drop falls
each time the tapper is triggered
Check RDE:
−
Connect reference electrode with WE
cable
−
Install RDE properly (see section 3.5)
−
Regenerate RDE (see section 3.5.2)
r/a