7.3 Error messages

746 VA Trace Analyzer / 747 VA Stand

7-27

Error message

Meaning/Causes

Rectification

Exit

SUBSTANCES - ERROR 3:
Param. Mass conc.
missing

Missing value for the

Mass

conc.

parameter

Enter

Mass conc.

parameter on the

”SUBSTANCES” page

Esc

SUBSTANCES - ERROR 4:
Param. Scope f./r.half
only with linear

The

f.half/r.half

pa-

rameter is admissible only with

a linear baseline

Change

Scope

or

Type

parameter on

the ”SUBSTANCES” page

Esc

SUBSTANCES - ERROR 5:
Param. Scope
f./r.double missing
counterpart

Missing counterpart for the

evaluation with

f.double

or

r.double

; in double peak

evaluation the first peak must

be evaluated with

f.double

and the second peak with

r.double

Check

Scope

parameter for both

substances and change

Esc

SUBSTANCES - ERROR 6:
Param. Scope
f./r.double direction
conflict

Wrong assignment of the two

peaks in the double peak

evaluation; the first peak (in

sweep direction) must be

evaluated with

f.double

, the

second with

r.double

Check

Scope

parameter for both

substances and change

Esc

SUBSTANCES - ERROR 7:
Undefined error

Non-specified error on the

”SUBSTANCES” page

Check the ”SUBSTANCES” page

Esc

VA STAND - ERROR 1:
Missing device

Connection to 747 VA Stand

broken:

Connecting cable not

plugged in

Wrong address

Hardware fault at VA Stand

Check connection to 747 VA Stand:

Plug in connecting cable

Correct address and restart 746 VA

Trace Analyzer

Run diagnostic routine for VA Stand

E/S

VA STAND - ERROR 2:
Working electrode
faulty

Fault at MME working electr.:

No connection

Too less mercury

Bad mercury quality

Dirty or clogged capillary

Dirty or damaged sealing

needle

No inert gas pressure

Dirty electrode body

Fault at reference electrode

No solution in measuring

vessel

No electrolyte in measuring

vessel

Tapping strength in DME

operation too weak

Fault at RDE working electrode:

No connection

Electrode tip not firmly

screwed on to the drive shaft

Contaminated electrode

surface

Check MME:

Connect reference electrode with WE

cable

Fill MME with mercury (see section

3.4.2 and 3.4.7)

Clean MME (see section 3.4.9) and

use ultrapure mercury (w

0.99999)

or filtered mercury

Change capillary (see section 3.4.8)

Clean MME (see section 3.4.9) and

replace sealing needle if need be

Check inert gas connection (see

section 3.3.3)

Rub the MME body dry, clean with

distilled water of dilute acid and dry

with a lint-free cloth

see

VA STAND - ERROR 4

Add measuring solution

Add a few drops of KCl 3 mol/L to the

measuring solution

Turn slotted screw 71

71 on the 747 VA

Stand during operation in an anti-

clockwise direction until a drop falls

each time the tapper is triggered

Check RDE:

Connect reference electrode with WE

cable

Install RDE properly (see section 3.5)

Regenerate RDE (see section 3.5.2)

r/a