3.3 Installation of 747 VA Stand

746 VA Trace Analyzer / 747 VA Stand

3-7

3 Attachment of a 747 VA Stand for the 695 Autosampler

Switch off 746 VA Trace Analyzer using the mains switch 11

11.

Plug one end of the third 6.2135.000 EBUS cable into connector of the

second EBUS cable and fasten together by tightening the screws in the

cable connector.

Plug the other end of the third EBUS cable into connection 34

34 of the third

747 VA Stand and fasten to the instrument by tightening the screws in the

cable connector.

Using a screwdriver, set stand address 35

35 of the 747 VA Stand to "3"

(arrow must point to the number "3").

3.3.2

Equipping the measuring head

The fixtures inserted in the openings and connections of the measuring head 41

41 in

the 747 VA Stand depend on the working electrode selected (MME or DME) (see

Fig. 9). The fully equipped measuring head for operation with a multi-mode elec-

trode is illustrated in section 2 (Figs 7 and 8), that for operation with a rotating disk

electrode in section 3.5 (Fig. 16).
When equipping the measuring head for the first time, the best procedure is as fol-

lows:

1 Preparations

Prepare multi-mode electrode MME 38

38 (details, see section 3.4) or rotat-

ing disk electrode RDE (details, see section 3.5) for operation.

Prepare reference electrode 43

43 (details, see section 3.6) for operation.

Tilt back cover 24

24 of measuring head arm.

2 Insert dummy stoppers

without automatic rinsing:

Screw dummy stopper 62

62 (6.1446.040) into opening 73

73.

Screw dummy stopper 61

61 (6.1446.040) into opening 74

74.

with and without automatic rinsing:

Screw dummy stopper 42

42 (6.1446.040) into opening 78

78.

Insert 6.2730.030 Stopper into enclosed nipple 44

44 and push enclosed

O-ring over the stopper.

Screw nipple 44

44 with stopper into opening 80

80 by hand.

3 Insert 4-way microtip

Remove stopper from nipple 44

44 and insert 4-way microtip 47

47 into nipple

44

44 as far as it will go.

Tighten nipple 44

44 using a 6.2739.010 Wrench until the 4-way microtip 47

47

can no longer move.

Pull the 4 lengths of PTFE tubing of the 4-way microtip 47

47 in succession

from above through the opening 86

86 (connection of 685 Dosimat, see

section 3.9; connection of 700 Dosino, see section 3.10).