4.2 Lead determination with standard addition
746 VA Trace Analyzer / 747 VA Stand
4-23
4.2.7
Storing the determination
The lead determination you have just performed remains in the working storage un-
til you start a new determination or load a different determination into the working
storage. For permanent storage of the recorded determination, you have the fol-
lowing three possibilities available:
1
Storage in determination memory
•
Press the yellow
MINATIONS" appears showing the determinations stored in the instru-
ment:
Files : 0 of 256 DETERMINATIONS
Available: 1353520 Bytes Full condition: stop
File
ДДДДДДДДДДДД
ДДДДДДДДДДДД ДДДДДДДДДДДД
ДДДДДДДДДДДД ДДДДДДДДДДДД
ДДДДДДДДДДДД ДДДДДДДДДДДД
ДДДДДДДДДДДД
Modify
column
Modify
line
Delete
file(s)
Copy
to
Copy
from
Unselect
file(s)
Change
display
>>
The header of the dialog page "DETERMINATIONS" contains information on the
number of stored determinations and the available storage space. This is followed by
the names of all stored determinations displayed in 4 columns.
You will find further information regarding this dialog page in section 5.4.2.
•
Press the softkey
. The softkey bar changes and presents
the sources for copying a determination to the determination memory.
•
Press the softkey
. The determination loaded in the
working storage is copied to the determination memory, the file name (in
this example
03201614.dtm
) appears in the first column under the title
File
.
•
Press the softkey
. The type of display changes, in
addition to the name of the determination a field for content information
and the storage space needed for the determination in bytes appear.
•
Both file name and content information can be changed for improved
documentation. To do this, press the softkey
to enter the
edit mode. The first 8 characters of the determination name can now be
changed in any way (the extension
.dtm
must be retained). Now press
the
Content
field for the
determination (e.g.
Lead determination with method TESTPb of
20.03.1997
). After confirmation with the
the navigation mode.