226 dos>a, O, s – Metrohm 746 VA Trace Analyzer User Manual
Page 377

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-226
DOS>A
Page:
O, S
Stand:
A
Abbreviation:
DO
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. DOS>A ..... Soln.name ........ V.add ........mL
Function
Dispense auxiliary solution into auxiliary vessel (AUX) on the 695
Autosampler.
Parameter
Range; Default value / Meaning
Soln.name
8 characters; -
Name of the auxiliary solution
The entry of a solution name is mandatory when the auxiliary solution
should be added automatically. In this case the solution name
Soln.name
must be entered in the parameter of the same name on the
"SOLUTIONS" page and assigned a position
Pos.
(see section 5.3.4).
Without entry of a name for
Soln.name
in the
DOS>A
instruction, without
entry of this name on the "SOLUTIONS" page or without a position
assignment, the auxiliary solution must be dispensed manually (the
prompt
DOS "Soln.name" "V.add" > A
then appears during the
program run).
V.add
0.010...50.000 mL; 1.000 mL
Solution volume that should be dispensed
Volume of the auxiliary solution that should be dispensed into the
auxiliary vessel.
Description
The
DOS>A
instruction is used for the manual or automatic addition of
auxiliary solutions (e.g. supporting electrolyte, reagent solution) to the
auxiliary vessel (AUX) on the 695 Autosampler.
For automatic addition with the 695 Autosampler, the appropriate
position of the auxiliary solution must be specified on the "SOLUTIONS"
page (see
Soln.name
).
Comments
The
DOS>A
instruction can be used only in operation with the 695
Autosampler.
see also
DOS/M, DOS>M, DOS>F, PDOS>M, SDOS>M, SDOS>A, SDOS>F