7.3 Error messages

746 VA Trace Analyzer / 747 VA Stand

7-31

-> test beep: y/n ? n

The instrument is now powered up and the ”MONITORING” page appears.

6

Load method or determination
Select the "METHODS", "DETERMINATIONS" or "DATA CARD" page and

copy desired method or determination to the working storage.

Overall initialization

Reinitialization of the entire hardware and software.
The instrument is reinitialized completely, the data loaded in the method and de-

termination storage as well as the method and determination loaded in the working

storage are cleared and all instrument parameters are reset to their default values.
The overall initialization is performed as follows:

1

Start reset
Initiate a reset using one of the 3 methods listed under Normal Reset, then

press the space key several times when the instrument is powering up until

the following screen display appears:

RAM initialization 746

ДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДД

-> contrast = ###

hit any key to confirm contrast

2

Adjust screen contrast
The screen contrast is now continuously varied from completely dark to

extremely light. Press any key as soon as the contrast has reached the

desired value. (If you are not satisfied with the result, you can improve the

contrast setting later on the ”VA TRACE ANALYZER” page.)

3

Run RAM test
Next you are asked whether or not you wish to perform a RAM test. If an

error has been found, it is advisable to press here. If the RAM test is

successful, the following screen display appears:

-> contrast = 108

-> non destructive RAM test: y/n ? y

- running non destructive RAM test

RAM test passed ok