5.5.3 CURVES
746 VA Trace Analyzer / 747 VA Stand
5-105
Substance-VR-Curve: XXXXXXXX-XX-smoothed
U.peak : XXXXXX mV dU.front: XXXXXX mV dU.rear: XXXXXX mV
I.peak : XXXXXX xA S.front: XXXXXX S.rear: XXXXXX
P.peak : XXXXXX xW
U.width: XXXXXX mV
Modify
column
New
display
Reset
param.
Save
param.
Softkeys
Meaning
Modify
column
Modify recognition parameters column-by-column
Modification of the parameters for the substance recognition. For
details of the recognition parameters, see section 5.6.3.
Modifications made here initiate a recalculation and curve display on
closure of the entry. These changes are temporary, they are canceled
with
or after exit with
The changes are permanently incorporated in the method only if they
are stored with
before exiting. To ensure they are also
effective for the result calculation, a recalculation must be initiated on
the "MONITORING" page with
.
New
display
Redisplay curve
This softkey is used for manual initiation of a recalculation and curve
display (this normally runs automatically each time the parameter
modification is closed).
Reset
param.
Reset recognition parameters
The parameters appearing in the window are reset to the values
entered on the "SUBSTANCES" page.
Save
param.
Store recognition parameters
The parameters appearing in the window are transferred to the
"SUBSTANCES" page and hence stored in the method.
10
Recognition
––––––––––––––––––––––––
U.verify ........mV
U.tol (+/-) ........mV
U.width min ........mV
U.width max ........mV
I.threshold ........pA
8