5.5.3 CURVES

746 VA Trace Analyzer / 747 VA Stand

5-105

Substance-VR-Curve: XXXXXXXX-XX-smoothed

U.peak : XXXXXX mV dU.front: XXXXXX mV dU.rear: XXXXXX mV

I.peak : XXXXXX xA S.front: XXXXXX S.rear: XXXXXX

P.peak : XXXXXX xW

U.width: XXXXXX mV

Modify

column

New

display

Reset

param.

Save

param.

Softkeys

Meaning

Modify

column

Modify recognition parameters column-by-column

Modification of the parameters for the substance recognition. For
details of the recognition parameters, see section 5.6.3.

Modifications made here initiate a recalculation and curve display on
closure of the entry. These changes are temporary, they are canceled
with

or after exit with .

The changes are permanently incorporated in the method only if they
are stored with

before exiting. To ensure they are also

effective for the result calculation, a recalculation must be initiated on
the "MONITORING" page with

.

New

display

Redisplay curve

This softkey is used for manual initiation of a recalculation and curve
display (this normally runs automatically each time the parameter
modification is closed).

Reset

param.

Reset recognition parameters

The parameters appearing in the window are reset to the values
entered on the "SUBSTANCES" page.

Save

param.

Store recognition parameters

The parameters appearing in the window are transferred to the
"SUBSTANCES" page and hence stored in the method.

10

Recognition

––––––––––––––––––––––––

U.verify ........mV

U.tol (+/-) ........mV

U.width min ........mV

U.width max ........mV

I.threshold ........pA

8

ESC