beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 511

background image

7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-12

Error message

Meaning/Causes

Rectification

Exit

DIAGNOSIS - ERROR 2:
Left some keys
unpressed

Diagnostic fault; not all keys

pressed in keypad test

Repeat keypad test

E/C

DIAGNOSIS - ERROR 3:
RS232 receive timeout

Diagnostic fault; error in RS232

test

Inform Metrohm service

r/a

DIAGNOSIS - ERROR 4:
RS232 sendtext
receivetext diff.

Diagnostic fault; error in RS232

test

Inform Metrohm service

E/C

DIAGNOSIS - ERROR 5:
RS232 config. register
1 read failed

Diagnostic fault; error in RS232

test

Inform Metrohm service

E/C

DIAGNOSIS - ERROR 6:
RS232 config. register
2 read failed

Diagnostic fault; error in RS232

test

Inform Metrohm service

E/C

DIAGNOSIS - ERROR 7:
Loopback, on-chip test
failed with code ....

Diagnostic fault; no 6.496.8480

Test Plug attached to ”RS 232

Interface #” output

Attach 6.496.8480 Test Plug to ”RS 232

Interface #” output and repeat test

E/C

DIAGNOSIS - ERROR 8:
Control lines, line
read failed

Diagnostic fault; no 3.496.8550

Remote Test Plug attached to

”Control Lines” output

Attach 3.496.8550 Remote Test Plug to

”Control Lines” output and repeat test

E/C

DIAGNOSIS - ERROR 9:
EBUS loopback connector
can not be accessed

Diagnostic fault; no EBUS test

plug attached to ”External Bus”

output

Attach EBSU test plug to ”External Bus”

output and repeat test

E/C

DIAGNOSIS - ERROR 10:
EBUS I/O lines read
failed

Diagnostic fault; error in EBUS

test

Inform Metrohm service

E/C

DIAGNOSIS - ERROR 11:
EBUS, detected request
not confirmed

Diagnostic fault; error in EBUS

test

Inform Metrohm service

E/C

DIAGNOSIS - ERROR 12:
EBUS, unexpected
request detected

Diagnostic fault; error in EBUS

test

Inform Metrohm service

E/C

DIAGNOSIS - ERROR 13:
EBUS, expected request
not received

Diagnostic fault; error in EBUS

test

Inform Metrohm service

E/C

DIAGNOSIS - ERROR 14:
Data card not found

Diagnostic fault; no data card

found

Insert data card

E/C

DIAGNOSIS - ERROR 15:
Data card battery stage
warning level ...

Diagnostic fault; battery run

down (low) or discharged (bad)

Change battery of data card

E/C

DIAGNOSIS - ERROR 16:
Data card battery word
bad

Diagnostic fault; test pattern in

RAM of data card destroyed,

i.e. the battery of the data card

is probably discharged

Change battery of data card

E/C

DIAGNOSIS - ERROR 17:
Data card not
initialized (formatted)

Diagnostic fault; non-formatted

data card

Format data card

E/C

DIAGNOSIS - ERROR 18:
Data card slot number
system corrupt

Diagnostic fault; error in testing

the data card

Inform Metrohm service

E/C

DIAGNOSIS - ERROR 19:
Data card unchained
block(s) marked as in
use

Diagnostic fault; error in ”Data

card confidence” test

Copy all files from data card to method

and/or determination memory, reformat

data card

E/C

DIAGNOSIS - ERROR 20:
Can’t test, while data
card is write protected

Diagnostic fault; data card is

write protected

Disable write protection of data card or

use a different data card

E/C