Page 128

background image

4.4 Lead determination with internal standard

746 VA Trace Analyzer / 747 VA Stand

4-43

5

Drinking water

Water that has been in the supply system for a lengthy period of time

absorbs metals from the pipe and fittings materials and thus falsifies the

analysis results. Hence, before taking the water sample allow the water

to run out of the tap in a powerful stream for several minutes.

The drinking water sample should be acidified immediately after the

sample has been taken by adding 10

µ

L conc. hydrochloric acid per

10 mL drinking water.

Owing to the large temperature dependence of the measured values (

2%/°C), we

recommend you to use the 6.1418.220 Thermostatable measuring vessel with

the internal standard method.

4.4.2

Loading the method

INTSTD.mth

To load the method

INTSTD.mth

from the method memory into the working stor-

age, the following steps are necessary:

1

Switch on instrument

Switch on the 746 VA Trace Analyzer with mains switch 11

11 at the rear of

the instrument. The instrument is initialized. If all is in order, the dialog

page "MONITORING" appears after a few seconds.

2

Switch to dialog page "METHODS"

Press the yellow key. The "METHODS" dialog page ap-

pears showing the methods stored in the instrument.

3

Selection of the method

INTSTD.mth

Use the cursor keys to select the method

INTSTD.mth

, which then

becomes black-backed.

4

Copying the method

INTSTD.mth to the working storage

Press the softkey

. The softkey bar changes and presents the

destinations for copying the selected method.

Press the softkey

. The method

INTSTD.mth

is

copied to the working storage.

4.4.3

Method specifications of the method

INTSTD.mth

In this section the 5 dialog pages of the group "Method Specifications" associated

with the loaded method

INTSTD.mth

will be explained in detail. If you are interested

only in a rapid lead measurement, you can proceed directly to section 4.4.4.

1

Switch to dialog page "OPERATION SEQUENCE"

Press the yellow key. The dialog page "OPERATION

SEQUENCE" appears with the main program for the lead determination: