5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-235
EXIT
Page:
O, S
Stand:
M, A
Abbreviation:
EX
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. EXIT .....
Function
Immediate exit from a loop or a block if the test condition is met.
Parameter
Range; Default value / Meaning
(Test)
EV:
STATE =,<> ctrl.line code; -
Test condition for decision
The following test conditions can be entered in the field under
Main
parameters
:
EV:
MC:
S0 < | <= | >= | > const. / C##
S1 < | <= | >= | > const. / C##
S2 < | <= | >= | > const. / C##
S3 < | <= | >= | > const. / C##
V0 < | <= | >= | > const. / C##
STATE = | <> ctrl.line code
The terms have the following meaning:
EV:
Evaluation value (
I.peak
,
I.wave,
P.peak
or
t.peak
) for the specified substance (
= substance name) in A, W or s
MC:
Mass concentration
ρ
s
for the specified sub-
stance (
=substance name) in g/L
S0
Sample size
S0
in g or L
S1,S2,S3
Sample specific variables
S1
,
S2
,
S3
V0
Sample volume in measuring vessel
V0.sample
in L
STATE
State of the input control lines 1...8 of the
remote interface
const.
Constant (numeric value)
C##
Common variable C01 ... C18 from the
working storage (see section 5.4.5)
ctrl.line code
8-digit numeric code for the state of the in-
put control lines 1...8. The following values
can be entered for each line:
0
off, inactive (5 V)
1
on, active (0 V)
*
immaterial