Metrohm tiamo 2.4 Manual User Manual
Page 991
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
975
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the mea-
suring point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
1.0 - 2000.0 mV/s
Default value
25.0 mV/s
Break point evaluation
Break point evaluation
on | off (Default value: off)
If this option is activated, then sharp, almost 90° changes in directions in
the measurement curve will be evaluated.
EP criterion
Measure of the minimum sharpness of the break point. The smaller the EP
criterion set, the more break points will be found. As this is a relative value
related to the total measured value modification, even small changes in
the measured value can be evaluated as a break point for a small mea-
sured value range.
Input range
0 - 1.0
Default value
0.3
Slope
Minimum difference between the slope before and after the break point.
The smaller the difference, the more break points will be found.
Input range
0.0 - 10.0
Default value
0.9
Smoothing factor
The higher the smoothing factor, the fewer endpoints will be found.
Input range
2 - 20
Default value
5