2 stirring, 1 stirring - overview, Stirring 3 – Metrohm tiamo 2.4 Manual User Manual
Page 1619
■■■■■■■■■■■■■■■■■■■■■■
8 Manual control
tiamo 2.4
■■■■■■■■
1603
device, it will automatically be reduced during dosing to the largest possi-
ble value.
NOTE
The filling rate should be reduced for viscous liquids.
Selection
maximum
Default value
maximum
Titrando, 814, 815, 855
Input range
0.01 - 166.00 mL/min
778, 789
Input range
0.01 - 160.00 mL/min
[Dosing]
Start manual dosing for the selected dosing device. Dosing will take place
as long as the button is pressed down. The dosed volume is displayed live.
NOTE
Parameters modified after the start of the dosing are not valid until the
next dosing procedure.
[Fill]
Start filling of the buret for the selected dosing device. During the filling
procedure the volume display is set to 0.0000 mL.
8.4.2
Stirring
8.4.2.1
Stirring - Overview
Program part: Manual control
If in the subwindow for the device selection (see Chapter 8.3, page 1597),
a Stirrer connected via MSB or stirrer connector is selected, then the
functions and parameters belonging to the stirrers are displayed in the
subwindow Functions/Parameters.
Stirrer functions
The stirrer functions are displayed on the following tabs:
■
■