10 set u - fixed endpoint evaluation, 5 set ipol, 1 set ipol - overview – Metrohm tiamo 2.4 Manual User Manual
Page 755
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
739
5.6.3.4.4.10
SET U - Fixed endpoint evaluation
Dialog window: Method
▶ SET U ▶ Properties... ▶ Additional evaluations ▶ Fixed
endpoint evaluation
▶ [New/[Properties] ▶ Fixed endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Input range
-2000.0 - 2000.0 mV
Time
Input range
0.0 - 999999.9 s
Volume
Input range
0.00000 - 9999.99 mL
5.6.3.4.5
SET Ipol
5.6.3.4.5.1
SET Ipol - Overview
Dialog window: Method
▶ SET Ipol ▶ Properties... ▶ SET Ipol - 'Command name'
Command for endpoint titrations with voltametric measurement
(selectable polarization current).
Devices
This command can be executed with the following instruments:
Titrando: 808, 809, 835, 836, 841, 842, 852, 857, 888, 901, 902, 904,
905, 906, 907
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798,
799
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
without conditioning
with conditioning