9 det ipol - measured value window – Metrohm tiamo 2.4 Manual User Manual
Page 552
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
536
■■■■■■■■
tiamo 2.4
5.6.3.2.4.9
DET Ipol - Measured value window
Dialog window: Method
▶ DET Ipol ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with measured value window (Ipol) ▶ [New/[Proper-
ties]
▶ Measured value window
Measured value windows are regions (windows) on the measured
value axis for which different parameters for potentiometric evaluation
can be defined. Only those endpoints which lie within these windows and
additionally meet the parameters defined for each window will be recog-
nized. Only one endpoint will be recognized per window.
Lower limit
Lower limit for the measured value window.
Titrando, 855
Input range
-2000.0 - 2000.0 mV
Default value
-2000.0 mV
Titrino
Input range
-2000 - 2000 mV
Default value
-2000 mV
Upper limit
Upper limit for the measured value window.
Titrando, 855
Input range
-2000.0 - 2000.0 mV
Default value
2000.0 mV
Titrino
Input range
-2000 - 2000 mV
Default value
2000 mV
EP criterion
Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.
Titrando, 855
Input range
0 - 200
Default value
5
EP recognition
Filter for the recognition of equivalence points: