7.14 Avantes spectrometer
■■■■■■■■■■■■■■■■■■■■■■
1560
■■■■■■■■
tiamo 2.4
Only editable if the analog output is switched on.
Input range
1 - 214,783,647 mAU/mV
Default value
100 mAU/mV
7.14.4
Properties - Detector
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Detector
The Detector tab shows information on the detector that is built into the
spectrometer.
Detector type
Shows the detector type.
Minimum wavelength
Shows the minimum wavelength of the measurable range in nm (instru-
ment-specific parameter).
Maximum wavelength
Shows the maximum wavelength of the measurable range in nm (instru-
ment-specific parameter).
Number of pixels
Shows the resolution with which the spectrum is recorded (instrument-
specific parameter).
7.14.5
Properties - Calibration
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Calibration
On the tab Calibration the data of the wavelength calibration is dis-
played. A specific wavelength, defined by four wavelength coefficients, is
assigned to each pixel on the detector. The wavelength calibration is car-
ried out with the command CAL Spec, the calibration data is calculated
automatically and displayed on this tab.
Date
Display of the time and date of the wavelength calibration.
User
Display of the user logged in at the time of the wavelength calibration.
Method
Display of the method the wavelength calibration has been carried out
with.