Metrohm tiamo 2.4 Manual User Manual
Page 927
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
911
Lower limit
Lower limit of the temperature. If the temperature falls below this limit,
the event Lower temperature limit violated is triggered.
Titrando, 855
Input range
–20.0 - 150.0 °C
Default value
–20.0 °C
Titrino
Input range
–170.0 - 500.0 °C
Default value
–170.0 °C
Lower hysteresis
Lower hysteresis of the temperature. If the temperature exceeds the lower
limit by this hysteresis value again, then the event Lower temperature
limit OK is triggered.
Titrando, 855
Input range
0.0 - 150.0 °C
Default value
0.2 °C
Upper limit
Upper limit of the temperature. If the temperature exceeds this limit, then
the event Upper temperature limit violated is triggered.
Titrando, 855
Input range
–20.0 - 150.0 °C
Default value
150.0 °C
Titrino
Input range
–170.0 - 500.0 °C
Default value
500.0 °C
Upper hysteresis
Upper hysteresis of the temperature. If the temperature once again falls
below the upper limit by this hysteresis value, then the event Upper tem-
perature limit OK is triggered.
Titrando, 855
Input range
0.0 - 150.0 °C
Default value
0.2 °C
Action
Selection of the event to be triggered if the upper or lower limit is viola-
ted: