2 stdadd man - general/hardware – Metrohm tiamo 2.4 Manual User Manual
Page 1089
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1073
Identifi-
cation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
not ever been started
.CONC
Concentration of the solution used for the command
.DBL
Total duration for the processing of the command in s
.EME
Final measured value (measured value after processing of
the command) in the unit of the measured value
.ENP
Electrode zero point calculated from the calibration
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has been ended at
least once, 0, invalid (variable not available) = Com-
mand has never been ended
.IME
Initial measured value (measured value before start condi-
tions are processed) in the unit of the measured value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.MTE
Temperature measurement with sensor; 1 = on, 0 = off)
.RES
Calculated result of the standard addition in the selected
unit
.SLO
Electrode slope calculated from the calibration
.STY
Type of stop with which the command was stopped: 1 =
normal;0 = manual or after error
.VAR
Variance of the calculated result of the standard addition
in the selected unit
.VOL
Dosed volume
5.6.4.17.2
STDADD man - General/Hardware
Tab: Method
▶ STDADD man ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the control device, the sensor and the stirrer
are defined here.