10 set ipol - fixed endpoint evaluation, 6 set upol, 1 set upol - overview – Metrohm tiamo 2.4 Manual User Manual
Page 777
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
761
[New]
Opens the External measured value # dialog window, in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 934).
[Properties]
Opens the External measured value # dialog window, in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 934).
[Delete]
Deletes the external measured value selected in the table.
5.6.3.4.5.10
SET Ipol - Fixed endpoint evaluation
Dialog window: Method
▶ SET Ipol ▶ Properties... ▶ Additional evalua-
tions
▶ Fixed endpoint evaluation ▶ [New/[Properties] ▶ Fixed endpoint evalua-
tion #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Input range
-2000.0 - 2000.0 mV
Time
Input range
0.0 - 999999.9 s
Volume
Input range
0.00000 - 9999.99 mL
5.6.3.4.6
SET Upol
5.6.3.4.6.1
SET Upol - Overview
Dialog window: Method
▶ SET Upol ▶ Properties... ▶ SET Upol - 'Command name'
Command for endpoint titrations with amperometric measurement
(selectable polarization voltage).
Devices
This command can be executed with the following instruments: