3 meas opt - measuring parameters – Metrohm tiamo 2.4 Manual User Manual
Page 1065
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1049
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.
Stirrer
The Stirrer section is only displayed for Device type = 089 Photometer.
Stirrer
Selection of the stirrer.
Selection
1 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate.
Input range
0 - 15
Default value
5
Switch off automatically
on | off (Default value: on)
If this check box is activated, the stirrer will be switched off automatically
when the command has finished.
5.6.4.14.3
MEAS Opt - Measuring parameters
Tab: Method
▶ MEAS Opt ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measuring parameters for the Avantes spectrometer
Wavelength
Wavelength at which the measured quantity (absorbance) is recorded.
Input range
100.0 - 2,000.0 nm
Default value
400.0 nm