Metrohm tiamo 2.4 Manual User Manual
Page 1152
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1136
■■■■■■■■
tiamo 2.4
from other measuring command) by which every measured absorbance
value can then be corrected.
Input range
–4,000 - 4,000 mAU
Default value
0 mAU
Integration time
Integration time for absorbance measurement.
Input range
0.01 - 600,000 ms
Default value
6 ms
Averaged spectra
Number of spectra that are recorded and averaged.
Input range
1 - 10,000
Default value
10
Smoothing
Number of neighboring pixels with which the light intensity of each mea-
surement pixel is determined.
Input range
1 - 100 pixels
Default value
1 pixels
Flash rate
Number of flashes per second of a pulsed xenon lamp as light source.
Input range
0 - 100 Hz
Selection
off
Default value
off
off
For the continuous type of the light source.
Measuring parameters for the 089 Photometer
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.
Input range
0 - 4,000 mAU
Default value
0 mAU
Measurement
Two types of measurement can be selected, Measurement with drift
control or Measurement without drift control.