Metrohm tiamo 2.4 Manual User Manual
Page 1026
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1010
■■■■■■■■
tiamo 2.4
Titrando: 808, 809, 835, 836, 857, 904, 905, 906, 907
pH/Ion - Meter: 867
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command MEAS Conc are configured on the fol-
lowing 3 tabs:
■
General/Hardware
Parameters for devices, sensors and stirrers.
■
Measuring parameters
Parameters for setting the measurement procedure.
■
Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.
Command variables
The following command variables are generated in the method run by the
command MEAS Conc and can be used in formulas under the designa-
tion 'Command name.Variable designation':
Designa-
tion
Description
.BLV
Blank value of the sensor used for the command (only
for ISE sensors)
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.DBL
Total duration for the processing of the command in s
.EME
End measured value (measured value after processing of
the command) in the unit of the measured value
.ENP
Electrode slope of the sensor used for the command (in
mV for ISE sensors)
.ETE
End temperature (temperature following processing of
the command) in °C