Metrohm tiamo 2.4 Manual User Manual
Page 899
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
883
Monitoring Dosing rate
Monitoring Dosing rate
on | off (Default value: off)
If this option is enabled, then the averaged dosing rate will be monitored
and any violated limits will be entered in the measuring point list. The
monitoring of the dosing rate is not started until 10 s after the start of the
titration.
Lower limit
Lower limit of the dosing rate. If the dosing rate falls below this limit, the
event Dosing rate lower limit violated is triggered.
Titrando, 855
Input range
0.00 - 166.00 mL/min
Default value
0.00 mL/min
Titrino
Input range
0.00 - 150.00 mL/min
Default value
0.00 mL/min
Lower hysteresis
Lower hysteresis of the dosing rate. If the dosing rate exceeds the lower
limit by this hysteresis value again, then the event Dosing rate lower
limit OK is triggered.
Titrando, 855
Input range
0.00 - 166.00 mL/min
Default value
0.20 mL/min
Upper limit
Upper limit of the dosing rate. If the dosing rate exceeds this limit, then
the event Dosing rate upper limit violated is triggered.
Titrando, 855
Input range
0.00 - 166.00 mL/min
Default value
166.00 mL/min
Titrino
Input range
0.00 - 150.00 mL/min
Default value
150.00 mL/min