5 Method
tiamo 2.4
■■■■■■■■
795
User
Editing the individual titration parameters which affect the titration
rate.
Control
NOTE
Will be displayed only if the user option is selected in the selection list
Titration rate.
Dynamics
The control range defines the measured value range before the specified
endpoint. It has a decisive influence on the titration rate and thus also on
the accuracy. Within the control range, dosing is controlled by the Min.
volume increment. The closer the endpoint, the slower the dosing until
the Min. volume increment has been reached. The larger the control
range, the slower the titration. Outside the control range, dosing is carried
out continuously at Max. rate.
Titrando
Input range
0.1 - 2,000.0 mV
Default value
300.0 (slow), 100.0 (optimal), 30.0 (fast) mV
Selection
off
off
The entire measured value range is defined as control range. The mini-
mum volume increment is applied throughout the entire titration.
Titrino
Input range
1 - 2,000 mV
Default value
300 (slow), 100 (optimal), 30 (fast) mV
Selection
off
off
The entire measured value range is defined as control range. The mini-
mum volume increment is applied throughout the entire titration.
Max. rate
Rate at which dosing is carried out outside of the Control range. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.