Metrohm tiamo 2.4 Manual User Manual
Page 828

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
812
■■■■■■■■
tiamo 2.4
890
Selection
1
Default value
1
Sensor
Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.
Selection
Sensor name | Metal electrode
Default value
Metal electrode
U(pol)
The polarization voltage is the voltage applied to the polarizable electrode
during an amperometric measurement.
Titrando
Input range
–1,250 - 1,250 mV (Increment: 25)
Default value
400 mV
Titrino
Input range
–1,270 - 1,270 mV (Increment: 10)
Default value
400 mV
Electrode check
on | off (Default value: off)
If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.
Temperature measurement
Type of temperature measurement.
Titrando
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
under Temperature on the Titration parameters tab will be used.