5 kft upol - titration parameters, Titration parameters – Metrohm tiamo 2.4 Manual User Manual
Page 835
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
819
Titrando
Input range
1 - 999 µL/min
Default value
10 µL/min
Delay time
If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time. The selection inf. stands for infin-
ite.
Titrando, 855
Input range
0 - 999 s
Default value
10 s
Titrino
Input range
0 - 999 s
Default value
10 s
Selection
inf.
Stop time
The titration is canceled when the absolute stop time defined here has
elapsed since the start of the titration. This parameter is displayed only if
Stop criterion = Time and if Delay time = inf..
Titrino
Input range
1 - 999,999 s
5.6.3.5.3.5
KFT Upol - Titration parameters
Tab: Method
▶ KFT Upol ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the running of the titration are defined on this tab.
Titration direction
Selection of the titration direction.
Selection
+ | - | auto
Default value
auto
+
positive measured value change
-
negative measured value change