6 set upol - stop conditions, Stop conditions – Metrohm tiamo 2.4 Manual User Manual
Page 792
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
776
■■■■■■■■
tiamo 2.4
temperature will be measured continuously. This value is used for temper-
ature correction in pH measurements.
Titrando, 855
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
Titrino
Input range
-170.0 - 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando, 855
Input range
0.1 - 999999 s
Default value
2.0 s
Titrino
Input range
0.08 - 16200 s
Default value
2.00 s
5.6.3.4.6.6
SET Upol - Stop conditions
Tab: Method
▶ SET Upol ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Entry
25 characters
Conditions for stopping the titration. If several stop conditions are set,
then the criterion which is fulfilled first will stop the titration.
Stop volume
Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Titrando, 855
Input range
0.00000 - 9999.99 mL
Default value
100,000 mL
Selection
off
Titrino
Input range
0.000 - 9999.99 mL
Default value
100.00 mL
Selection
off