4 meas conc - additional measured values – Metrohm tiamo 2.4 Manual User Manual
Page 1032
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1016
■■■■■■■■
tiamo 2.4
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Input range
0 - 1E+99
Selection
off
Default value
off
off
No stop.
Temperature
Temperature
Manually entered titration temperature. If a temperature sensor is con-
nected and the Temperature measurement is set on the General/
Hardware tab under Sensor to automatic or continuous, then the
temperature will be measured continuously.
Input range
-20.0 - 150 °C
Default value
25.0
5.6.4.9.4
MEAS Conc - Additional measured values
Tab: Method
▶ Commands ▶ MEAS Conc ▶ Properties... ▶ Additional measured
values
Command name
Name of the command.
Entry
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together with the measured values present in
the default settings in additional measured value columns Calculated #
or External # in the measuring point list.
Additional calculated measured values
Additional calculated measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be
presented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.
[New]
Open the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be entered (see Chapter
5.6.3.11.1, page 933).