MET Cond - Start conditions
Tab: Method
▶ MET Cond ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Entry
25 characters
The start conditions are processed in the sequence shown before the con-
ductivity measurement is started.
Initial measured value
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:
Signal drift
The measured value is only accepted if the signal drift is lower than the
value entered here.
Input range
0.001 - 999.0 (mS/cm)/min
Selection
off
Default value
off
off
The measured values will be accepted when the maximum waiting
time has elapsed.
Min. waiting time
The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.
Input range
0 - 999999 s
Default value
0 s
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.
Input range
0 - 999999 s
Default value
1 s