10 kfc - fixed endpoint evaluation, 7 brc, 1 brc - overview – Metrohm tiamo 2.4 Manual User Manual
Page 860
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
844
■■■■■■■■
tiamo 2.4
5.6.3.6.10
KFC - Fixed endpoint evaluation
Dialog window: Method
▶ KFC ▶ Properties... ▶ Additional evaluations ▶ Fixed
endpoint evaluation
▶ [New]/[Property] ▶ Fixed endpoint evaluation #
Measured quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Water
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Input range
-2000.0 - 2000.0 mV
Time
Input range
0.0 - 999999.9 s
Water
Input range
0.0 - 999999.9 µg
5.6.3.7
BRC
5.6.3.7.1
BRC - Overview
Dialog window: Method
▶ BRC ▶ Properties... ▶ BRC - 'Command name'
Command for coulometric titration for determining the bromine con-
sumption. Out of the amount of bromine consumed the bromine index
(amount of bromine consumed (in mg) per 100 g sample) and the bro-
mine number (amount of bromine (in g) per 100 g sample) can be calcu-
lated.
Instruments
This command can be executed with the following devices.
Titrando: 851, 852
Appearance
The command has the following appearance: