2 met cond - general/hardware, General/hardware – Metrohm tiamo 2.4 Manual User Manual
Page 693
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
677
Identification
Description
.SVT
Total start volume (volume that was added accord-
ing to all three start conditions) in mL
.TC.TC
Temperature coefficient in %/°C
.TC.CO
Coefficient c0 of the Chebyshev polynomial of the
solution used
.TC.C1
Coefficient c1 of the Chebyshev polynomial of the
solution used
.TC.C2
Coefficient c2 of the Chebyshev polynomial of the
solution used
.TC.C3
Coefficient c3 of the Chebyshev polynomial of the
solution used
.TC.C4
Coefficient c4 of the Chebyshev polynomial of the
solution used
.TC.MAX
Maximum temperature coefficient in %/°C
.TC.MIN
Minimum temperature coefficient in %/°C
.TC.TSTART
Start temperature in °C
.TC.TSTOP
Stop temperature in °C
5.6.3.3.6.2
MET Cond - General/Hardware
Tab: Method
▶ MET Cond ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the instrument, the dosing device, the sensor
and the stirrer are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.