Metrohm tiamo 2.4 Manual User Manual
Page 1021
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1005
Identifica-
tion
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has not ever been started
.DBL
Total duration for the processing of the command in s
.EGF
Last measured gas flow rate (measured value following
processing of the command) in mL/min
.EME
Final measured value (measured value after processing
of the command) in the unit of the measured value
.FIN
Command status; 1 = Command has been ended
at least once, 0, invalid (variable not available) =
Command has never been ended
.GMA
Maximum gas flow rate in mL/min
.GMI
Minimum gas flow rate in mL/min
.GMN
Average gas flow rate in mL/min
.IGF
Initial gas flow (measured value at the time of the start
of the command) in mL/min
.IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
.LP.CAx
Calculated value × (1...3) for the last measuring point
of the measuring point list
.LP.EXx
External value × (1...3) for the last measuring point of
the measuring point list
.LP.GFL
Gas flow value for the last measuring point of the
measuring point list in mL/min
.LP.MEA
Measured value (temperature) for the last measuring
point of the measuring point list in the unit of the
measured value
.LP.TEM
Temperature for the last measuring point of the mea-
suring point list in °C
.LP.TIM
Time in s until the last measuring point of the measur-
ing point list is reached
.MA.MEA
Maximum measured value (temperature) in the unit of
the measured value