7 set u - conditioning, Conditioning – Metrohm tiamo 2.4 Manual User Manual
Page 749
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
733
Stop time
Stops when the preset time (including start conditions) has elapsed since
the start of the titration.
Titrando, 855
Input range
0 - 999999 s
Selection
off
Default value
off
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.
Titrando, 855
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Input range
0.01 - 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.4.4.7
SET U - Conditioning
Tab: Method
▶ SET U ▶ Properties... ▶ Conditioning
Command name
Name of the command.
Entry
25 characters
Conditioning
on | off (Default value: off)
Enable/disable conditioning. If conditioning is disabled, the following
parameters are not visible.
Start drift
The determination can only be started when the currently measured drift
is less than the entered start drift entered here. If this is the case, then
the status displayed in the Run subwindow in the Workplace program
part changes to COND READY.