5 cal spec - calibration wavelength, 6 cal loop conc, 1 cal loop conc - overview – Metrohm tiamo 2.4 Manual User Manual
Page 1129: Cal loop conc
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4
■■■■■■■■
1113
The reference wavelengths with which the spectrometer is calibrated can
be entered on this tab.
Tolerance
Tolerance for the preset wavelength when assigning the highest peak
maximum found during the calibration measurement to the reference
wavelength. If a peak is outside the range Wavelength +/- Tolerance,
the calibration will be canceled. The existing calibration in the configura-
tion remains in effect.
Input range
0.1 - 1,100.0 nm
Default value
0.5 nm
[New]
Opens the Calibration wavelength # dialog window for entering wave-
lengths and tolerances (see Chapter 5.6.5.5.5, page 1113).
[Properties]
Opens the Calibration wavelength # dialog window to edit the values
selected in the table (see Chapter 5.6.5.5.5, page 1113).
[Delete]
Deletes the selected line.
5.6.5.5.5
CAL Spec - Calibration wavelength #
Tab: Method
▶ CAL Spec ▶ Properties... ▶ Calibration wavelengths ▶ [New/[Prop-
erties]
▶ Calibration wavelength #
Calibration wavelengths
Input range
200.0 - 1,100.0 nm
Default value
'empty' nm
5.6.5.6
CAL LOOP Conc
5.6.5.6.1
CAL LOOP Conc - Overview
Dialog window: Method
▶ CAL LOOP Conc ▶ Properties... ▶ CAL LOOP Conc -
'Command name'
Loop command for the calibration of ion-selective electrodes, used
together with the CAL MEAS Conc command.
Appearance
The command has the following appearance: