tiamo 2.4
5.6.3.7.5
BRC - Titration parameters
Tab: Method
▶ BRC ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the running of the titration are defined on this tab.
Extraction time
The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, e.g. for the titration of sparingly soluble samples.
Input range
0 - 999999 s
Default value
0 s
Temperature
Manually entered titration temperature.
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Input range
0.1 - 999999.0 s
Default value
2.0 s
5.6.3.7.6
BRC - Stop conditions
Tab: Method
▶ BRC ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Entry
25 characters
The conditions for stopping the titration are defined on this tab.
Stop time
Stop when the entered time has elapsed since the start of the titration.
off means no stop.
Input range
1 - 999999 s
Selection
off
Default value
off