2 cal meas conc - general/hardware – Metrohm tiamo 2.4 Manual User Manual
Page 1134
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1118
■■■■■■■■
tiamo 2.4
Identifica-
tion
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has not ever been started
.DBL
Total duration for the processing of the command in s
.EME
Final measured value (measured value after processing
of the command) in the unit of the measured value
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has been ended at
least once, 0, invalid (variable not available) = Com-
mand has never been ended
.IME
Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.MEA
Measured value for the last measuring point in the mea-
suring point list in the unit of the measured value
.LP.TEM
Temperature for the last measuring point of the measur-
ing point list in °C
.LP.TIM
Time in s until the last measuring point in the measuring
point list is reached
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in the measuring point list
.STY
Type of stop with which the command was stopped: 1
= normal;0 = manual or after error
5.6.5.7.2
CAL MEAS Conc - General/Hardware
Tab: Method
▶ CAL MEAS Conc ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the device, the sensor and the stirrer are
defined on this tab.